Cancer Informatics Collaboratory
Home / Datasets / FDA benchmark index
Index · Medical device CFD · US FDA

FDA CFD Benchmarks: Index

Start here to find each FDA benchmark, the papers behind it and where to download the files today.

The research behind this page

  • Prasanna Hariharan, PhD: Mechanical engineer at the US FDA Center for Devices and Radiological Health since 2008; PhD, University of Cincinnati. ✓ FDA speaker profile

These researchers created the original work. They did not write or review this page. ✓ marks a credential or role confirmed on an official profile; "as listed" roles come from the original page. All researchers featured on this site.

Based on: 4 peer-reviewed papers, 1 government source, 1 official project or publisher site, 2 archived original pages. See all sources

Get the files

Benchmark 1
Nozzle: PIV velocity and pressure at Re 500–6,500
Benchmark 2
Centrifugal blood pump: PIV, hemolysis, H–Q
Download
GitHub repository with both benchmarks
Geometry paper (nozzle)
Instructions and CAD files were published with a 2012 paper in Cardiovascular Engineering and Technology
Links checked 21 Nov 2026.

Pages in this set

Benchmark 1 in brief

A two-way nozzle in which the diameter changes gradually (a cone) at one end and suddenly at the other. Three laboratories measured velocity with PIV using a blood-analog fluid matched to the nozzle's refractive index, at throat Reynolds numbers of 500, 2,000, 3,500, 5,000 and 6,500 (laminar to turbulent).[1]

Who uses them

By the end of 2021 the working group had listed 35 outside studies using the benchmarks, from university groups to software makers.[2]

Key publications

  1. 2012Stewart SFC, et al. Assessment of CFD performance in simulations of an idealized medical device: results of FDA's first computational interlaboratory study. Cardiovasc Eng Technol. 2012.
    DOI: 10.1007/s13239-012-0087-5
  2. 2011Hariharan P, et al. Multilaboratory particle image velocimetry analysis of the FDA benchmark nozzle model to support validation of computational fluid dynamics simulations. J Biomech Eng. 2011.
    DOI: 10.1115/1.4003440 · PubMed 21428676
  3. 2017Malinauskas RA, et al. FDA benchmark medical device flow models for CFD validation. ASAIO J. 2017.
    DOI: 10.1097/MAT.0000000000000499 · PubMed 28114192

Identifiers checked against PubMed and the publishers' DOI records, 22 Nov 2026.

Sources

  1. Original nozzle page (archived copy) · web.archive.org · Archived original
  2. Citation list (archived copy) · web.archive.org · Archived original

About this page

No copy of this exact address was archived. We built this index from the archived nozzle pages of the same FDA wiki (main page, publications and citations).