FDA Benchmark Centrifugal Blood Pump
A simplified centrifugal blood pump that FDA built, tested and released so that anyone can check a pump CFD model against real measurements.
The research behind this page
- Prasanna Hariharan, PhD: Mechanical engineer at the US FDA Center for Devices and Radiological Health since 2008; PhD, University of Cincinnati. ✓ FDA speaker profile
These researchers created the original work. They did not write or review this page. ✓ marks a credential or role confirmed on an official profile; "as listed" roles come from the original page. All researchers featured on this site.
Based on: 2 peer-reviewed papers, 1 government source, 1 official project or publisher site, 2 archived original pages. See all sources
Get the data
- Publisher
- US FDA, Center for Devices and Radiological Health
- Device
- Simplified centrifugal blood pump: a rotor in a pump chamber
- What's measured
- Flow fields by PIV, blood damage (hemolysis), pump head vs. flow (H–Q)
- Files
- Instructions package with CAD geometry and flow conditions; experimental data
- Where it's hosted
- GitHub (OSEL-DAM), with FDA's comparison report as a PDF
What was built and measured
FDA engineers made acrylic copies of the pump so they could see inside it. They mapped the flow with particle image velocimetry (PIV), measured blood damage in separate experiments, and recorded the pump's performance as a head–flow (H–Q) curve.
The instructions package gives everything needed to set up a simulation: the CAD geometry of the chamber and rotor, and the flow rates and rotor speeds to run.[1]
What the 24-participant comparison found
FDA ran an interlaboratory study in which 24 participants simulated the pump and compared their results with the measurements. The spread between groups was wide, especially under transient conditions. The practical lesson: one good match at a single operating point is not enough to call a model validated.[2]
Who has used it
Software makers and research groups have used the pump as a public test case. Examples include a peer-reviewed modelling study of the pump and a series of COMSOL application posts in several languages.
Key publications
- 2020Good BC, et al. Computational modeling of the Food and Drug Administration's benchmark centrifugal blood pump. Artif Organs. 2020.
DOI: 10.1111/aor.13643 · PubMed 31971269 · Free full text (PMC) - 2017Malinauskas RA, et al. FDA benchmark medical device flow models for CFD validation. ASAIO J. 2017.
DOI: 10.1097/MAT.0000000000000499 · PubMed 28114192
Identifiers checked against PubMed and the publishers' DOI records, 22 Nov 2026.
Sources
- Original pump page (archived copy) · web.archive.org · Archived original
- FDA comparison report (PDF) · www.fda.gov · Government source
- The working group's citation list (archived copy) · web.archive.org · Archived original
About this page
This page was the benchmark #2 page of FDA's CFD/Blood Damage project wiki on the hub, written by Prasanna Hariharan (created Mar 2017, last edited May 2022). We rewrote it around the data and where it lives now.
- Archived copy, 18 Jan 2023: ncihub.cancer.gov/wiki/FDA_CFD/ComputationalRoundRobin2Pump
- Live sources last checked: 21 Nov 2026